- Last 7 days
-
corona-323.quarto.pub corona-323.quarto.pub
-
through SEM
SEM cannot give you quantitative height information.
-
5x5 μm
should be area!
-
-
corona-323.quarto.pub corona-323.quarto.pub
-
Abstract
Date missing.
-
- Oct 2024
-
camille2341.quarto.pub camille2341.quarto.pub
-
123, 96, 244, 159, and 124 degrees
observe the numbers on the angles are multiples of 30
-
people who took the time to make the samples provided by
Yes, I should have mentioned that.
-
Filip
need full name!
-
properties
general, be specific
-
SmartScan
need to flatten image; caption not quite matches the content, it is not a PtTe2 image, it is an image of a PtTe2 flake on top of a substrate.
-
found to be 835.3 nm with an
Looking at the fit; you can see that it is a very poor fit, so the correlation length must be much less known. It is the average particle size, but here you only have one particle (flake), so the error would be quite large.
-
nanoAfm
nanoAFMr is the correct spelling; needs citation!
-
igure 3: Image
This is not the topography, but the error image.
-
2024913_PHYS545_JCCBAB_AFM_T020200122Si16
not needed in the report; again in the notebook, but in the report those details are "hidden" in the code portion.
-
T020200122Si16
Sample name is good for notebooks; however, in the report, we rename the sample usually to something simpler, such as the gold sample, etc.
-
seconds
s, also use 333.2 kHz (note the spacing)
-
(H)
Hurst value $H$
-
0.37 nm
1 significant digits only
-
flake
PtTe should not be italics.
-